The detection limit (LOD) is the lowest concentration a method can reliably detect — that is, distinguish from blank — and corresponds to qualitative detectability; the quantitation limit (LOQ) is the lowest concentration that can be quantified with the required precision and accuracy, usually about three times the LOD or as defined by the method. Together they are the core sensitivity indicators of method validation or verification, determined by both the method principle and the instrument conditions, and should be comparable across laboratories.
Use it during method validation or verification in chromatography, spectroscopy, environmental and pharmaceutical laboratories, when writing validation reports under ICH Q2, ISO/IEC 17025 or industry standards, and when checking that sensitivity meets regulatory limits. Choose the algorithm by the data available: calibration curve, signal-to-noise, or blank measurements.
Enter the calibration data (concentration versus response), choose the algorithm, and read the LOD and LOQ. Then prepare verification solutions near the LOD and LOQ and measure them repeatedly to confirm recovery and precision — for example RSD at the LOQ within 10–20% and a signal-to-noise ratio of at least 3 at the LOD. Record the verification results in the method validation report; if they deviate from the calculation, matrix interference is likely and sample preparation or conditions need optimization.
Calibration-curve method: LOD = 3.3σ/S and LOQ = 10σ/S, where σ is the standard deviation of blank or lowest-concentration responses (or the regression residual SD) and S is the slope. Signal-to-noise method: LOD at 3×S/N and LOQ at 10×S/N. Blank method: LOD = mean_blank + 3×SD_blank and LOQ = mean_blank + 10×SD_blank. The 3.3 and 10 factors trace to ICH Q2: a signal-to-noise of 3 is detectable (one-sided α ≈ 0.05) and 10 is quantifiable (RSD about 10%).